Monte Carlo Study of the Scattering Error of a Quartz Reflective Absorption Tube Public Deposited

http://ir.library.oregonstate.edu/concern/defaults/s7526j39m

Descriptions

Attribute NameValues
Creator
Abstract or Summary
  • A Monte Carlo model was used to study the scattering error of an absorption meter with a divergent light beam and a limited acceptance angle of the receiver. Reflections at both ends of the tube were taken into account. Calculations of the effect of varying optical properties of water as well as the receiver geometry were performed. A weighting function showing the scattering error quantitatively as a function of angle was introduced. Some cases of practical interest are discussed.
Resource Type
DOI
Date Available
Date Issued
Citation
  • Piskozub, Jacek, Piotr J. Flatau, J. V. Ronald Zaneveld, 2001: Monte Carlo Study of the Scattering Error of a Quartz Reflective Absorption Tube. Journal of Atmospheric and Oceanic Technology, 18, 438–445.
Rights Statement
Publisher
Replaces
Additional Information
  • Vol. 18 no. 3.
  • description.provenance : Made available in DSpace on 2012-02-20T22:55:15Z (GMT). No. of bitstreams: 1 ZaneveldJVRonald.CEOAS.MonteCarloStudy.pdf: 107664 bytes, checksum: e29dd67d44c295020938c57a2ab57d7c (MD5) Previous issue date: 2001-03