  • A high-temperature, high-pressure autoclave has been adapted for use with a high-field nuclear magnetic resonance spectrometer. The autoclave has an internal pressure range of atmospheric pressure to 1.5 kbar and a temperature range of 273 K to 1900 K. The autoclave is usable in a high field (8 T) magnet with a room temperature, 76.4 mm bore. The autoclave was tested using assorted nuclear species with resonant frequencies ranging from 57 to 70 MHz at pressures ranging from atmospheric pressure to 1220 bar and temperatures ranging from 273 K to 448 K. Previously, the autoclave was used in conjunction with an iron magnet at temperatures to 1900 K and pressures to 1.5 kbar.
Resource Type
Date Available
Date Issued
Degree Level
Degree Name
Degree Field
Degree Grantor
Commencement Year
Advisor
Committee Member
Academic Affiliation
Non-Academic Affiliation
Subject
Rights Statement
Publisher
Peer Reviewed
Language
Digitization Specifications
  • File scanned at 300 ppi (Monochrome, 8-bit Grayscale, 24-bit Color) using ScandAll PRO 1.8.1 on a Fi-6670 in PDF format. CVista PdfCompressor 4.0 was used for pdf compression and textual OCR.