  • The three-grating optical interferometer is studied to identify restrictions on alignment and improve stability. A description of the way a three-grating interferometer works is given as well as a method for proper set up. The overall power loss through the three gratings is measured and can be used to estimate the amount of atoms that will be detected at the output of the atomic interferometer. Criteria are developed for misalignment affects. Equations are presented for spacing and rotation limitations which can then be applied to the atomic interferometer. A stabilization technique is developed using a servo system. The elimination of low frequency inertial noise is accomplished.