  • A Heath EUW-20A Servo-Recorder was converted to a self-balancing, recording Wheatstone bridge. The EUW-20A 10000Ω slide-wire was used as the variable resistance arm of the bridge, the servo-amplifier was used as the bridge unbalance signal detector, and the servo-motor was used to balance the bridge automatically. The bridge was arranged to produce a readout directly proportional to the conductance of the measured element. The problems accompanying the measurement of electrolytic conductance with this instrument were identified. There are two major sets of problems: Those arising from adverse cell conditions involving stirring rate, electrolyte concentration and electrode condition, and those arising from cell capacitance associated with the ionic double-layer at the electrodes. The cell capacitance causes a phase difference between the a-c potentials of two of the bridge arms. This phase difference causes a significant dead band in the recorder response. Two sets of conductometric titrations of Na₂CO₃ with HC1 were performed, one using a sine wave to excite the bridge and one using a square wave. The error in the results obtained from these titrations was generally less than one percent and the reproducibility was one-half percent or less. The instrument performed satisfactorily when the conductance change occurring during the titration was as low as five percent.
Resource Type
Date Available
Date Issued
Degree Level
Degree Name
Degree Field
Degree Grantor
Commencement Year
Advisor
Academic Affiliation
Non-Academic Affiliation
Subject
Dichiarazione dei diritti
Publisher
Peer Reviewed
Language
Digitization Specifications
  • File scanned at 300 ppi using Capture Perfect 3.0 on a Canon DR-9050C in PDF format. CVista PdfCompressor 5.0 was used for pdf compression and textual OCR.
Replaces

Le relazioni

Parents:

This work has no parents.

In Collection:

Elementi