Implementation of radiation film dosimetry system to be used for the verification of a 3-D electron pencil-beam algorithm on a radiation treatment planning system

Pubblico Deposited

Contenuto scaricabile

Scarica il pdf
https://ir.library.oregonstate.edu/concern/graduate_thesis_or_dissertations/th83m182w

Descriptions

Attribute NameValues
Creator
Abstract
  • Radiation film dosimetry process using the RIT 113 v.4 dosimetry software and the film digitizer VXR-l2plus was used to evaluate the accuracy of electron dose calculations of the RAHD radiation therapy treatment planning system at Samaritan Regional Cancer Center. Kodak Ready-Pack EDR-2 film is recommended for dose distribution analysis in clinically practical dose ranges. The pencil-beam algorithm has a limitation for calculating dose in the penumbra region and in the tail region where the dose falls off.